Patient radiation doses in various fluoroscopically guided orthopaedic procedures.

Citation:

Tsapaki V, Tsalafoutas IA, Fagkrezos D, Lazaretos I, Nikolaou VS, Efstathopoulos N. Patient radiation doses in various fluoroscopically guided orthopaedic procedures. Radiat Prot Dosimetry. 2016;168(1):72-5.

Abstract:

All orthopaedic fluoroscopic procedures performed using C-arm guidance were monitored for 1 y. The type of procedure, fluoroscopy time (T), kerma-area product (KAP) values and number of radiographs (F) were recorded. The two most often performed techniques were as follows: intramedullary nailing (IMN) of intertrochanteric/peritrochanteric (IP) fractures (101 cases, 49.3 %) and antergrade IMN of femur or tibia shaft (TS) fractures (28 cases, 13.7 %). For the remaining procedures, none accounted for >5 %, categorised as 'various' (76 cases, 37 %). Large variations in T, KAP and F were observed. For IMN of IP fractures, antergrade IMN of femur and TS fractures and for various procedures, respectively, median values were T--2.1, 2.2 and 0.6 min, KAP--6.3, 6.3 and 0.6 Gy cm(-2) and F--21, 2.2 and 6.7. The patient doses during fluoroscopically guided procedures are relatively low compared with other interventional procedures.