AUTOMATED FLOW-INJECTION TECHNIQUE FOR USE IN DISSOLUTION STUDIES OF SUSTAINED-RELEASE FORMULATIONS - APPLICATION TO IRON(II) FORMULATIONS

Citation:

Georgiou CA, Valsami GN, Macheras PE, Koupparis MA. AUTOMATED FLOW-INJECTION TECHNIQUE FOR USE IN DISSOLUTION STUDIES OF SUSTAINED-RELEASE FORMULATIONS - APPLICATION TO IRON(II) FORMULATIONS. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS. 1994;12:635-641.

Abstract:

The application of flow-injection analysis (FIA) to automated dissolution studies of sustained-release formulations is described. The long-term stability of the dissolution-FIA analyser was checked during unattended operation for 42 h. The construction of multiple calibration curves with the so-called electronic dilution FIA procedure was used to extend the linear range of the determination. The computer-controlled FIA system and the principles of associated software are described and applied to dissolution studies of sustained-release formulations of iron(II) using its sensitive reaction with the colour reagent, ferrozine. The extended linear range of the determination is 1-130 ppm iron(II) and the precision (RSD) better than 3% (n = 3).